Semi Auto Biochemistry Analyzer

Analytical Modes:kinetic,fixed time,end-point,absorbance

  • FOB Price
    NA
  • Min Order Quantity
    NA
  • Payment Terms
    NA
  • Share:
Request for Quote

Other Details

Features

    Analytical Modes:kinetic,fixed time,end-point,absorbance

    Wavelengths:7 standard filters:340,405,492,510,546,578,630nm,1 more filter optional

    Full open reagent system

    Items can programmed,number of items not limited

    Voice prompting,making the operation interesting

    Display:LCD with high brightness

    Input:Key board

    Real time graph display and print

    Lamp saving feature,longer lamp life

    Large storage ability,up to1000,000 results

Technical Specifications

    Assay Method: reaction rate, endpoint, two-point, double wavelength, coefficient, multiple standards, blank reagent, blank sample, absorbance at fixed point, Q.C, Q.C graph and turbidity

    Temperature System: PELTIER temperature control component, 25,37,accuracy:

    Light Source: 6V, 10W halogen lamp of more than 2000 hours life time

    GML Precision:2nm

    FWHM: 10nm

    Stray Light:0.5%(absorbance 2.5)

    Absorbance Range:-0.214~3.000

    Absorbance Linearity: 1%

    Absorbance Repeatability: 0.001

    Precision: 1%

    Carry Over: 1%

    Resolution: 1m Abs

    Stability: 3mAbs/h

    Flow Cell:permanent quartz flow cell,minimal volume is 32uL,and disposable flow cell is selectable

    Sampling System:pressure auto releasing pumps ensure the accuracy of suction volume

    Absorption Volume:200uL~3000uL

    Sample Volume:10uL~50uL

    Display:240*128LCD

    Interface:RS-232 serial port

    Working Condition:Temperature:10-30,Relative Humidity:20%-80%

    Voltage:AC 85~264V,auto adapt according to the input voltage

    Power: 100W

    Quality System:CE