Benchtop laser micrometer

Measurement system and display integrated into a single unit for easybench-top inspection.• A statistical calculation function is provided.• Included standard RS-232C interface and SPC output interface. Technical DataRange: 0.5 - 60

  • FOB Price
    NA
  • Min Order Quantity
    NA
  • Payment Terms
    NA
  • Share:
Request for Quote

Other Details

Measurement system and display integrated into a single unit for easybench-top inspection.• A statistical calculation function is provided.• Included standard RS-232C interface and SPC output interface. Technical DataRange: 0.5 - 60mmResolution: 0.00005 - 0.1mm(.000002” - .005”) ( ): Inch/Metric modelsLinearity at 20°C*: ±2.5μmRepeatability (±2σ): ±0.6μmPositional error**: ±2.5μmMeasuring region: 10 x 60mmLaser wavelength: 650nm, VisibleNumber of scans: 1600 scans/sLaser scanning speed: 226m/sApplicable standard for laser: IEC (FDA: 544-116-1A),Main display: 16-digit fluorescent tubeInterface units equipped: RS-232C, SPC, Foot switchPower supply: 100 - 240V AC±10%, 40VA, 50/60Hz* At the center of the measuring region.** Potential measurement error due to variation in workpiece position within themeasuring region.FunctionZero-setting, Presetting, GO/±NG judgment, Multi-limitjudgment, Offset value setting, Sample measurement,Statistical calculation, Group judgment, Data output,Workpiece position display, mm/inch switching, Dual-gagecalibration, Transparent object measurement, Automaticmeasurement, Abnormal data elimination